User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: General Radio 1864:Megomhmeter DATE: 04-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 337 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 34401A STANDARD: DTB Megohm Resistance Box ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 STD DTB Megohm Resistance Box 1.004 HEAD INITIAL CONDITIONS 1.005 DISP Plug in the UUT and turn on the power. 1.005 DISP Allow the UUT to warm-up for one-half hour. 1.006 HEAD {* Voltage Accuracy *} 1.007 JMP 2.002 1.008 EVAL 2.001 ACC 10.000V 0.00041U 2.002 DISP [32] Connect the UUT as follows: 2.002 DISP + UNKNOWN .......................... 34401A Hi 2.002 DISP - UNKNOWN .......................... 34401A Lo 2.003 DISP [32] Set the UUT as follows: 2.003 DISP Test Voltage ....................... 1-0-V 2.003 DISP Discharge/Charge/Measure ........... Charge 2.004 IEEE [@34401] *RST 2.005 IEEE [@34401]CONF:VOLT:DC 2.006 IEEE [@34401][D2000]READ?[I] 2.007 MATH MEM1 = 10 2.008 MEME 2.009 MEMC V 2.000% #! Test Tol 0.2, Sys Tol 0.00041, TUR 487.805 (>= 4.00). 3.001 DISP [32] Set the UUT as follows: 3.001 DISP Test Voltage ....................... 2-0-V 3.002 ACC 20.000V 0.00162U 3.003 IEEE [@34401]CONF:VOLT:DC 3.004 IEEE [@34401][D2000]READ?[I] 3.005 MATH MEM1 = 20 3.006 MEME 3.007 MEMC V 2.00% #! Test Tol 0.4, Sys Tol 0.00162, TUR 246.914 (>= 4.00). 4.001 DISP [32] Set the UUT as follows: 4.001 DISP Test Voltage ....................... 3-0-V 4.002 ACC 30.000V 0.00207U 4.003 IEEE [@34401]CONF:VOLT:DC 4.004 IEEE [@34401][D2000]READ?[I] 4.005 MATH MEM1 = 30 4.006 MEME 4.007 MEMC V 2.00% #! Test Tol 0.6, Sys Tol 0.00207, TUR 289.855 (>= 4.00). 5.001 DISP [32] Set the UUT as follows: 5.001 DISP Test Voltage ....................... 4-0-V 5.002 ACC 40.000V 0.00252U 5.003 IEEE [@34401]CONF:VOLT:DC 5.004 IEEE [@34401][D2000]READ?[I] 5.005 MATH MEM1 = 40 5.006 MEME 5.007 MEMC V 2.00% #! Test Tol 0.8, Sys Tol 0.00252, TUR 317.460 (>= 4.00). 6.001 DISP [32] Set the UUT as follows: 6.001 DISP Test Voltage ....................... 5-0-V 6.002 ACC 50.000V 0.00297U 6.003 IEEE [@34401]CONF:VOLT:DC 6.004 IEEE [@34401][D2000]READ?[I] 6.005 MATH MEM1 = 50 6.006 MEME 6.007 MEMC V 2.00% #! Test Tol 1, Sys Tol 0.00297, TUR 336.700 (>= 4.00). 7.001 DISP [32] Set the UUT as follows: 7.001 DISP Test Voltage ....................... 6-0-V 7.002 ACC 60.000V 0.00342U 7.003 IEEE [@34401]CONF:VOLT:DC 7.004 IEEE [@34401][D2000]READ?[I] 7.005 MATH MEM1 = 60 7.006 MEME 7.007 MEMC V 2.00% #! Test Tol 1.2, Sys Tol 0.00342, TUR 350.877 (>= 4.00). 8.001 DISP [32] Set the UUT as follows: 8.001 DISP Test Voltage ....................... 7-0-V 8.002 ACC 70.000V 0.00387U 8.003 IEEE [@34401]CONF:VOLT:DC 8.004 IEEE [@34401][D2000]READ?[I] 8.005 MATH MEM1 = 70 8.006 MEME 8.007 MEMC V 2.00% #! Test Tol 1.4, Sys Tol 0.00387, TUR 361.757 (>= 4.00). 9.001 DISP [32] Set the UUT as follows: 9.001 DISP Test Voltage ....................... 8-0-V 9.002 ACC 80.000V 0.00432U 9.003 IEEE [@34401]CONF:VOLT:DC 9.004 IEEE [@34401][D2000]READ?[I] 9.005 MATH MEM1 = 80 9.006 MEME 9.007 MEMC V 2.00% #! Test Tol 1.6, Sys Tol 0.00432, TUR 370.370 (>= 4.00). 10.001 DISP [32] Set the UUT as follows: 10.001 DISP Test Voltage ....................... 9-0-V 10.002 ACC 90.000V 0.00477U 10.003 IEEE [@34401]CONF:VOLT:DC 10.004 IEEE [@34401][D2000]READ?[I] 10.005 MATH MEM1 = 90 10.006 MEME 10.007 MEMC V 2.00% #! Test Tol 1.8, Sys Tol 0.00477, TUR 377.358 (>= 4.00). 11.001 DISP [32] Set the UUT as follows: 11.001 DISP Test Voltage ....................... 10-0-V 11.002 ACC 100.00V 0.00522U 11.003 IEEE [@34401]CONF:VOLT:DC 11.004 IEEE [@34401][D2000]READ?[I] 11.005 MATH MEM1 = 100 11.006 MEME 11.007 MEMC V 2.00% #! Test Tol 2, Sys Tol 0.00522, TUR 383.142 (>= 4.00). 12.001 HEAD {} 12.002 JMP 13.001 12.003 EVAL 13.001 DISP [32] Set the UUT as follows: 13.001 DISP Test Voltage ....................... 1-1-V 13.002 ACC 11.000V 0.000445U #13.003 IEEE [@1281] DCV 10 #13.003 IEEE [@1281] [D1000]RDG?[I] 13.003 IEEE [@34401]CONF:VOLT:DC 13.004 IEEE [@34401][D2000]READ?[I] 13.005 MATH MEM1 = 11 13.006 MEME 13.007 MEMC V 2.00% #! Test Tol 0.22, Sys Tol 0.000445, TUR 494.382 (>= 4.00). 14.001 DISP [32] Set the UUT as follows: 14.001 DISP Test Voltage ....................... 1-2-V 14.002 ACC 12.000V 0.00048U 14.003 IEEE [@34401]CONF:VOLT:DC 14.004 IEEE [@34401][D2000]READ?[I] 14.005 MATH MEM1 = 12 14.006 MEME 14.007 MEMC V 2.00% #! Test Tol 0.24, Sys Tol 0.00048, TUR 500.000 (>= 4.00). 15.001 DISP [32] Set the UUT as follows: 15.001 DISP Test Voltage ....................... 1-3-V 15.002 ACC 13.000V 0.000515U 15.003 IEEE [@34401]CONF:VOLT:DC 15.004 IEEE [@34401][D2000]READ?[I] 15.005 MATH MEM1 = 13 15.006 MEME 15.007 MEMC V 2.00% #! Test Tol 0.26, Sys Tol 0.000515, TUR 504.854 (>= 4.00). 16.001 DISP [32] Set the UUT as follows: 16.001 DISP Test Voltage ....................... 1-4-V 16.002 ACC 14.000V 0.00055U 16.003 IEEE [@34401]CONF:VOLT:DC 16.004 IEEE [@34401][D2000]READ?[I] 16.005 MATH MEM1 = 14 16.006 MEME 16.007 MEMC V 2.00% #! Test Tol 0.28, Sys Tol 0.00055, TUR 509.091 (>= 4.00). 17.001 DISP [32] Set the UUT as follows: 17.001 DISP Test Voltage ....................... 1-5-V 17.002 ACC 15.000V 0.000585U 17.003 IEEE [@34401]CONF:VOLT:DC 17.004 IEEE [@34401][D2000]READ?[I] 17.005 MATH MEM1 = 15 17.006 MEME 17.007 MEMC V 2.00% #! Test Tol 0.3, Sys Tol 0.000585, TUR 512.821 (>= 4.00). 18.001 DISP [32] Set the UUT as follows: 18.001 DISP Test Voltage ....................... 1-6-V 18.002 ACC 16.000V 0.00062U 18.003 IEEE [@34401]CONF:VOLT:DC 18.004 IEEE [@34401][D2000]READ?[I] 18.005 MATH MEM1 = 16 18.006 MEME 18.007 MEMC V 2.00% #! Test Tol 0.32, Sys Tol 0.00062, TUR 516.129 (>= 4.00). 19.001 DISP [32] Set the UUT as follows: 19.001 DISP Test Voltage ....................... 1-7-V 19.002 ACC 17.000V 0.000655U 19.003 IEEE [@34401]CONF:VOLT:DC 19.004 IEEE [@34401][D2000]READ?[I] 19.005 MATH MEM1 = 17 19.006 MEME 19.007 MEMC V 2.00% #! Test Tol 0.34, Sys Tol 0.000655, TUR 519.084 (>= 4.00). 20.001 DISP [32] Set the UUT as follows: 20.001 DISP Test Voltage ....................... 1-8-V 20.002 ACC 18.000V 0.00069U 20.003 IEEE [@34401]CONF:VOLT:DC 20.004 IEEE [@34401][D2000]READ?[I] 20.005 MATH MEM1 = 18 20.006 MEME 20.007 MEMC V 2.00% #! Test Tol 0.36, Sys Tol 0.00069, TUR 521.739 (>= 4.00). 21.001 DISP [32] Set the UUT as follows: 21.001 DISP Test Voltage ....................... 1-9-V 21.002 ACC 19.000V 0.000725U 21.003 IEEE [@34401]CONF:VOLT:DC 21.004 IEEE [@34401][D2000]READ?[I] 21.005 MATH MEM1 = 19 21.006 MEME 21.007 MEMC V 2.00% #! Test Tol 0.38, Sys Tol 0.000725, TUR 524.138 (>= 4.00). 22.001 HEAD {} 22.002 JMP 23.001 22.003 EVAL 23.001 DISP [32] Set the UUT as follows: 23.001 DISP Test Voltage ....................... 10-0-0V 23.002 IEEE [@34401] CONF:VOLT:DC 1000, .001V 23.003 IEEE [@34401] VOLT:DC:RANG 1000 23.004 IEEE [@34401][D2000]READ?[I] 23.005 ACC 1000.00V 0.057U 23.006 MATH MEM1 = 1000 23.007 MEME 23.008 MEMC V 2.00% #! Test Tol 20, Sys Tol 0.057, TUR 350.877 (>= 4.00). 24.001 DISP [32] Set the UUT as follows: 24.001 DISP Discharge/Charge/Measure ........... Discharge 24.002 HEAD {* Resistance Tests *} 24.003 JMP 25.001 24.004 EVAL 25.001 DISP Disconnect the test setup. 25.001 DISP On the 5700A, disconnect the GUARD to GND link. 25.002 DISP [32] Set the UUT as follows: 25.002 DISP Test Voltage ....................... 1-0-V 25.002 DISP Discharge/Charge/Measure ........... Measure 25.002 DISP Range Switch ....................... 100k 25.003 DISP With the UUT UNKNOWN terminals open, adjust the 25.003 DISP SET INFINITY dial for an reading of INFINITY 25.003 DISP on the UUT meter. 25.004 DISP [32] Set the UUT as follows: 25.004 DISP Discharge/Charge/Measure ........... Discharge 25.005 DISP Connect the UUT ± UNKNOWN terminals to the Fluke 5700A 25.005 DISP OUTPUT V/ohm/A HI/LO terminals. 25.006 DISP [32] Set the UUT as follows: 25.006 DISP Discharge/Charge/Measure ........... Measure 25.007 HEAD {Test Voltage = 10} 25.008 JMP 26.001 25.009 EVAL 26.001 5700 100.000kZ S 2W 26.002 MEMI Enter the UUT meter reading. 26.003 MEM* 100 26.004 MEME 26.005 MEMC kZ 4.00% #! Test Tol 4000, Sys Tol 1.3, TUR 3076.923 (>= 4.00). 27.001 DISP [32] Set the UUT as follows: 27.001 DISP Range Switch ....................... 1M 27.002 5700 1.000MZ S 2W 27.003 MEMI Enter the UUT meter reading. 27.004 MEME 27.005 MEMC MZ 4.000% #! Test Tol 40000, Sys Tol 18, TUR 2222.222 (>= 4.00). 28.001 DISP [32] Set the UUT as follows: 28.001 DISP Range Switch ....................... 10M 28.002 5700 10.00MZ S 2W 28.003 MEMI Enter the UUT meter reading. 28.004 MEM* 10 28.005 MEME 28.006 MEMC MZ 4.000% #! Test Tol 4e+005, Sys Tol 370, TUR 1081.081 (>= 4.00). 29.001 MEME 29.002 DISP [32] Set the UUT as follows: 29.002 DISP Range Switch ....................... 100M 29.003 5700 100.0MZ S 2W 29.004 MEMI Enter the UUT meter reading. 29.005 MEM* 100 29.006 MEME 29.007 MEMC MZ 4.000% #! Test Tol 4e+006, Sys Tol 12000, TUR 333.333 (>= 4.00). 30.001 HEAD {} 30.002 HEAD {Test Voltage = 100} 30.003 JMP 31.001 30.004 EVAL 31.001 DISP [32] Set the UUT as follows: 31.001 DISP Discharge/Charge/Measure ........... Discharge 31.002 DISP Disconnect the UUT from the 5700A and connect it 31.002 DISP to the 1G ohm resistor on the DTB MEGOHM resistance box. 31.003 DISP [32] Set the UUT as follows: 31.003 DISP Test Voltage ....................... 1-0-0V 31.003 DISP Discharge/Charge/Measure ........... Charge 31.003 DISP Range Switch ....................... 1G 31.004 ACC 1.0000GZ 2% 31.005 MEMI Enter the calibrated value for the 1G resistor. 31.006 MEME 31.007 DISP [32] Set the UUT as follows: 31.007 DISP Discharge/Charge/Measure ........... Measure 31.008 MEMI Enter the UUT meter reading. 31.009 MEME 31.010 MEMC GZ 4.0000% #! WARNING: Test Tol 0.04, Sys Tol 0.02, TUR 2.000 (< 4.00). 32.001 DISP [32] Set the UUT as follows: 32.001 DISP Discharge/Charge/Measure ........... Discharge 32.001 DISP Range Switch ....................... 10G 32.001 DISP 32.002 DISP Connect the UUT to 10G ohm resistor 32.003 DISP [32] Set the UUT as follows: 32.003 DISP Discharge/Charge/Measure ........... Charge 32.004 ACC 10.0000GZ 2% 32.005 MEMI Enter the calibrated value for the 10G resistor. 32.006 MEME 32.007 DISP [32] Set the UUT as follows: 32.007 DISP Discharge/Charge/Measure ........... Measure 32.008 MEMI Enter the UUT meter reading. 32.009 MEM* 10 32.010 MEME 32.011 MEMC GZ 4.0000% #! WARNING: Test Tol 0.4, Sys Tol 0.2, TUR 2.000 (< 4.00). #33.001 HEAD {} #33.002 HEAD {Test Voltage = 500} 33.001 JMP 34.001 33.002 EVAL 34.001 DISP [32] Set the UUT as follows: 34.001 DISP Discharge/Charge/Measure ........... Discharge 34.002 DISP Test Voltage ....................... 5-0-0V 34.003 DISP Range Switch ....................... 100G 34.004 DISP Connect the UUT to 100G ohm resistor 34.005 DISP [32] Set the UUT as follows: 34.005 DISP Discharge/Charge/Measure ........... Charge 34.006 ACC 100.000GZ 2% 34.007 MEMI Enter the calibrated value for the 100G resistor. 34.008 MEME 34.009 DISP [32] Set the UUT as follows: 34.009 DISP Discharge/Charge/Measure ........... Measure 34.010 MEMI Enter the UUT meter reading. 34.011 MEM* 100 34.012 MEME 34.013 MEMC GZ 4.0000% #! WARNING: Test Tol 4, Sys Tol 2, TUR 2.000 (< 4.00). 35.001 DISP [32] Set the UUT as follows: 35.001 DISP Discharge/Charge/Measure ........... Discharge 35.002 END #! T.U.R.s less than 4.00: 3 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.